Recall Z-3242-2018 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Terumo 100/120V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801763

DateAug 28, 2018
StatusTerminated
ReasonScrews on a Terumo Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system.
Root causeProcess control
ActionTerumo CVS issued letter on 8/2/818 via Fed'X 2 day air to inform users of the problem for the Power Manager boards with unverifiable or out of specification screws, potential hazard and field correction. A Terumo Field Service Representative will contact users to schedule the field correction activities. Terumo CVS Customer Service: 1.800.521.2818, Monday Friday, 8 a.m. 6 p.m. ET
Quantity47 units
DistributionNationwide Foreign: Mexico, BELGIUM, CANADA, India, Indonesia, Singapore, South Korea , Vietnam