ClearanceBrief
Recall Z-0188-2019 ยท OGO Intraocular Pressure Lowering Implant

Alcon Research, LTD.

CyPass System 241, Intraocular Pressure Lowering Implant, CYPASS MICRO-STENT, CYPASS APPLIER, REF: 8065754010. For ophthalmic use.

DateOct 23, 2018 (initiated Aug 28, 2018)
StatusTerminated
ReasonAnalysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.
Root causeDevice Design
ActionThe firm initiated their market withdrawal by letter on 08/28/2018. The firm requested surgeons immediately cease implantation of the device and return any unused product.
Quantity480 units
DistributionWorldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Finland, Germany, Hong Kong, Iceland, Italy, Mexico, Peru, Singapore, Spain, Sweden, Switzerland, United Kingdom, and Urugua
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.