Recall Z-0069-2019 ยท FMG Stopcock, I.V. Set

ICU Medical, Inc.

ICU Medical 14" (36 cm) Appx 3.6 ml, Bifuse Transfer Set w/Rotating Luer, REF: 14237-28 The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).

DateAug 20, 2018
StatusCompleted
ReasonDesign change implemented changing the device from a blood set with a filter to a transfer set without a filter. Both devices had the same listing number. Customers may not be aware of the product design change and use the transfer set as a blood transfusion set without realizing a blood filter is not included.
Root causeDevice Design
ActionOn 08/20/18, the firm, icumedical, mailed "URGENT MEDICAL DEVICE RECALL" l notices to consignees. The notice described the product, problem and actions to be taken. If product has been further distributed, the firm requested consignees to notify users. Customers were advised to return affected product only if the blood filter is needed. Customers needing response forms and/or return labels were advised to call 1-800-479-0551. For further inquiries call: 1-866-829-9025.
Quantity2361 Units
DistributionU.S. Distribution in states of: NY, AL, CA, MS, PA, GA, MI, NC, WI, IL, WA, UT and FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K964435PRIMARY IV SET
Icu Medical, Inc., Apr 2, 1997