Recall Z-0063-2019 ยท DWC Controller, Temperature, Cardiopulmonary Bypass

Cincinnati Sub-Zero Products LLC, a Gentherm Company

Hemotherm 400CE, Catalog Numbers 86022 & 86023 The Hemotherm Dual Reservoir Cooler/Heater is used to lower, maintain, or raise the temperature of the water flowing through a Blood Oxygenator/Heat Exchanger that is used to cool or warm blood during cardiopulmonary bypass procedures lasting six hours or less. The Hemotherm Dual Reservoir Cooler/Heater may also be used with a hyper/hypothermia bla

DateOct 3, 2018 (initiated Sep 4, 2018)
StatusTerminated
ReasonPotential for membrane (control panel) switch errors leading to an inability of the device to change control settings or an unintended change to control settings.
Root causeDevice Design
ActionOn September 4, 2018, the firm, CSZ, issued an "Urgent Medical Device Correction" notices to customers overnight via courier service. Customers were advised to take the following actions: - Immediately examine your inventory and quarantine product subject to the field action. - If you choose to use the product prior to the correction, please consider having a backup device available. - In addition, if you have further distributed this product, please identify your customers and notify them at once of this medical device correction. Your notification to your customers may be enhanced by in
Quantity270 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122813HEMOTHERM MODEL 400CE COOLER/HEATER SYSTEM
Cincinnati Sub-Zero Products, Inc., Dec 18, 2012