Recall Z-3211-2018 ยท IYE Accelerator, Linear, Medical

Accuray Incorporated

Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systems

DateSep 20, 2018 (initiated Aug 31, 2018)
StatusTerminated
Reasonthe Ball Cube II phantom calibration film was not manufactured within necessary dimensions and may introduce up to 0.5 mm of positional inaccuracy to the CyberKnife System
Root causeNonconforming Material/Component
ActionThe firm, Accuray, disseminated their "Urgent Field Safety Notices" dated 08/31/2018 by email and letter to its customers. The letter described the product, problem and actions to be taken. The letter requested the customers cease use and quarantine the product for recovery and replacement by an Accuray Representative and complete and return the Acknowledgement Form via email at: astsauver@accuray.com or mail to: Accuray Incorporated, one Erdman place, Madison, WI 53719. If you have any questions, contact Vice President, Regulatory Affairs at 608-824-2800 or email: dbiank@accuray.com or th
Quantity259 units
DistributionWorldwide Distribution: United States (nationwide), and countries of: United Kingdom, Germany, France, Finland, Netherlands, Portugal, Hong Kong, Poland, Australia, United Arab Emirates, Thailand, and Japan.