ClearanceBrief
Recall Z-3310-2018 ยท EOQ Bronchoscope (Flexible Or Rigid)

Olympus Corporation of the Americas

Rhino Laryngo scope ENF-VT2

DateAug 8, 2018
StatusTerminated
ReasonThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Root causeOther
ActionUrgent Medical Device Safety Notice letters dated 8/8/18 were distributed to customers. The letters instructed customers to perform the following: Our records indicate that your facility has purchased one or more of the above-referenced endoscopes. OLYMPUS requires that you take the following actions: Inspect your inventory for the referenced Olympus models. Please keep this customer letter in your inventory and conduct the following action prior to every use. Step 1: Wearing sterile gloves, attempt to rotate or turn the instrument channel port in counter-clock direction as shown in pict
Quantity3701
DistributionThe products were distributed US nationwide and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061313EVIS EXERA 180 SYSTEM
Olympus Medical Systems Corporation, Aug 31, 2006