ClearanceBrief
Recall Z-3302-2018 ยท EOQ Bronchoscope (Flexible Or Rigid)

Olympus Corporation of the Americas

Bronchovideoscope BF-Q180-AC

DateAug 8, 2018
StatusTerminated
ReasonThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Root causeOther
ActionUrgent Medical Device Safety Notice letters dated 8/8/18 were distributed to customers. The letters instructed customers to perform the following: Our records indicate that your facility has purchased one or more of the above-referenced endoscopes. OLYMPUS requires that you take the following actions: Inspect your inventory for the referenced Olympus models. Please keep this customer letter in your inventory and conduct the following action prior to every use. Step 1: Wearing sterile gloves, attempt to rotate or turn the instrument channel port in counter-clock direction as shown in pict
Quantity655
DistributionThe products were distributed US nationwide and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050220EVIS EXERA BRONCHOVIDEOSCOPE, OLYMPUS XBF-160Y3AC, XBF-1T160Y3AC,XBF-Q160Y2AC
Olympus Medical Systems Corporation, Feb 10, 2005