ClearanceBrief
Recall Z-3284-2018 ยท EOQ Bronchoscope (Flexible Or Rigid)

Olympus Corporation of the Americas

Bronchofiberscope F-1T160

DateAug 8, 2018
StatusTerminated
ReasonThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Root causeOther
ActionUrgent Medical Device Safety Notice letters dated 8/8/18 were distributed to customers. The letters instructed customers to perform the following: Our records indicate that your facility has purchased one or more of the above-referenced endoscopes. OLYMPUS requires that you take the following actions: Inspect your inventory for the referenced Olympus models. Please keep this customer letter in your inventory and conduct the following action prior to every use. Step 1: Wearing sterile gloves, attempt to rotate or turn the instrument channel port in counter-clock direction as shown in pict
Quantity3317
DistributionThe products were distributed US nationwide and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023984MODIFICATION TO OLYMPUS BRONCHOSCOPES BF-40 SERIES, BF-240 SERIES, AND BF-160 SERIES
The Olympus Optical Co., Jan 31, 2003