Recall Z-0064-2019 ยท KPA Slide Stainer, Automated

Leica Microsystems, Inc.

Leica Biosystems HistoCore SPECTRA ST, Slide Stainer, Automated Product Usage: The intended use of this device is a stainer for laboratory applications and intended for the preparation of histological and cytological specimens.

DateJun 13, 2018
StatusTerminated
ReasonThese devices have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could lead to mains power on the metal plate resulting in hazardous voltage to the user.
Root causeDevice Design
ActionOn approximately 06/13/2018 sent a Field Action Notices to consignees. Instructions included to make arrangements with their Leica Service Engineer to correct affected devices, ensure all appropriate personnel are made aware of the recall, and complete and return the Field Action Notice Return Response Form to confirm receipt and understanding of the recall notice.
Quantity67 units
DistributionDistributed nationwide to AZ, CA, CO, KS, MA, MH, MO, NJ, OH, WI. Distributed worldwide to CANADA, CHINA, GERMANY, GREAT BRITAIN, JAPAN, SWITZERLAND
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.