Recall Z-3192-2018 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Invivo Corporation

Xper Flex Cardio Physiomonitoring System Model # 453564669081 Software Version 1.6.0.0388

DateAug 3, 2018
StatusTerminated
ReasonReleased with an incorrect, older firmware version installed. May result in intermittent loss of ECG monitoring, inaccurate ECG amplitude display and or auxiliary output, and inaccurate heart rate (HR) displayed due to QRS detection fault.
Root causeSoftware Manufacturing/Software Deployment
ActionPhilips is asking customers to follow the Action to be Taken by Customer/User section of the August 3, 2018, Field Safety Notice: The Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. Philips will replace all affected devices. Users should discontinue use of affected devices immediately. A Philips representative will contact you regarding your affected device.
Quantity38
DistributionDistribution in the Netherlands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101571XPER INFORMATION MANAGEMENT AND FLEX CARDIO PHYSIOMONITORING SYSTEMS
Witt Biomedical Corp, Oct 26, 2010