ClearanceBrief
Recall Z-3194-2018 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Fresenius Medical Care Renal Therapies Group, LLC

2008T HEMODIALYSIS SYS, WITH CDX, Part Number 190713, Applicable only with 2008T BlueStar Premium Upgrade Kit, Part Number 191138 Product Usage: Indicated for acute and chronic dialysis therapy.

DateJul 13, 2018
StatusTerminated
ReasonIn the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and a mandatory rinse should be required. However, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature enabled, the machine enters the dialysis program even though the dialysis program button is disabled and the mandatory rinse has not been performed. As a result, disinfecting chemical may not be completely removed from the machine before sta
Root causeUnder Investigation by firm
ActionOn 07/13/2018 phone calls were immediately conducted to the affected clinics. Contact was made to the Technical Service Directors explaining issue and to disable the Auto Start Function on their BlueStar SW v2.71 installed 2008T Hemodialysis Machines.
Quantity80
DistributionUS Nationwide in the states of : CA, CT, MI, NC, and NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K1739722008T BlueStar Hemodialysis Machine
Fresenius Medical Care Renal Therapies Group, LLC, May 10, 2018