ClearanceBrief
Recall Z-3240-2018 ยท KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component

Stryker GmbH

Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002

DateJul 26, 2018
StatusTerminated
ReasonWhen creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction plan for the axial portion of the plan.
Root causeUnder Investigation by firm
ActionStryker contracted Third Party: Stericycle Contact Center, 6026 Lakeside Blvd., Indianapolis, IN 46278 notified affected hospitals via 2-day UPS (with return receipt) on July 26, 2018 All affected surgeon users notified via email with Request read Receipt and Request delivered Receipt options selected. Provided the following: if you have selected the Correct Axial First button, with multiple corrections per day, redo the correction plan by either of the following methods: a. Calculate the correction plan with one correction per day selected on the Correction Plan page or b. If more than one
Quantity233
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163656Hoffmann LRF System
Stryker GmbH, Apr 14, 2017