Recall Z-0066-2019 ยท OOE Ophthalmic Femtosecond Laser
Johnson & Johnson Surgical Vision Inc
OptiMedica Catalys Precision Laser System, CE 0044 containing software cOS 5.00.33 Product Label for Serial Numbers: 44109015; 44106215; 44107315; and 44065613 label states "Manufactured for: OptiMedica Corporation 1310 Moffett Park Drive Sunnyvale, CA 94089 USA Product Label for Serial Numbers: 46120916; 46123116, 46124617; 46131418, and 46131618 label states "Manufactured for: OptiMedica Cor
| Date | May 9, 2018 |
| Status | Terminated |
| Reason | Software upgrade exhibits failures when executing Daily Alignment Verification (DAV), specifically, cuts on the plastic hemisphere during DAV do not match the intended cuts shown on the treatment overlay screen on the system GUI. |
| Root cause | Device Design |
| Action | On May 9, 2018, affected customers in the U.S. and Puerto Rico were contacted by phone and verbally informed to stop using the affected system. On May 22, 2018, the firm, Johnson & Johnson Vision, sent an "URGENT MEDICAL DEVICE FIELD REMOVAL" letter dated May 18, 2018 via Federal Express overnight delivery with signature required to all customers with a system containing the affected software installed. The customers were instructed to: STOP Using the affected system and complete and return the Customer Reply Form via fax to: Johnson & Johnson Vision Quality Assurance at 1-714-247-4510, ema |
| Quantity | 9 |
| Distribution | US Distribution to states of: CO, GA, IL, MD including Puerto Rico and internationally to countries of: Austria and Germany. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K113479 | OPTIMEDICA CATALYS PRECISION LASER SYSTEM Optimedica Corporation, Dec 21, 2011 |
| K121091 | CATALYS PRECISION LASER SYSTEM Optimedica Corporation, Aug 28, 2012 |
| K141079 | OPTIMEDICA CATALYS PRECISION LASER SYSTEM Optimedica Corporation, Sep 11, 2014 |
| K170322 | Catalys Precision Laser System Abbott Medical Optics, Inc., May 19, 2017 |
| K172002 | Catalys Precision Laser System Amo Manufacturing USA, LLC, Aug 25, 2017 |