Recall Z-3022-2018 ยท MCW Catheter, Peripheral, Atherectomy

Micro Therapeutics Inc, Dba Ev3 Neurovascular

ev3 TurboHawk PERPHERAL PLAQUE EXCISION SYSTEM FOR SURGICAL USE, LARGE VESSEL XTENDED TIP FOR CALCIUM, Model THS-LX-C for peripheral vascular use.

DateAug 10, 2018
StatusTerminated
ReasonThe Distal Flush Tool (DFT) accessories that were packaged in this lot number are not compatible with the TurboHawk LX-C catheter. This may result in the inability to remove excised material from the tip of the catheter.
Root causeProcess control
ActionMedtronic began notifying the Consignees on 10-Aug-2018. Consignees were asked to quarantine and return affected product.
Quantity67 units
DistributionDistribution to US states of KS, CA, FL, GA, AZ, MI, LA, TX, NE, KY, and China