Diagnostica Stago, Inc.

Diagnostica Stago STA Noplastine CI Plus, 10 ml vial. In-Vitro Diagnostic for determination of the prothrombin time (PT) in plasma. (REF 00667) The STA - Noplastine CI Plus kits provide reagents for the determination of the prothrombin time (PT) in plasma with STA-R, STA Compact and STA Satellite. (In the USA this procedure has been assigned to the moderate complexity category per CLIA 1988 - C

DateJul 24, 2018
StatusTerminated
ReasonQC values outside of the assigned ranges (prolonged Prothrombin time / decreased PT %)
Root causeDevice Design
ActionDiagnostica Stago issued notification on July 24, 2018 to US Customers provided the reason for the notification, actions to be taken to mitigate the risk. Action: If you have in your laboratory, any of the lots among those listed in the appendix, we are asking you to: If it is not already done, run a Quality Control test at every change of vial. Return to your Stago affiliate, by fax or e-mail, the enclosed form completed and confirming that you have read this letter. Questions contact Stago Hotline (1-800-725-0607)
Quantity24043 kits
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall