Recall Z-3378-2018 ยท BSO Catheter, Conduction, Anesthetic

B. Braun Medical, Inc.

CE18TKY 18GA SAFETY TUOHY NEEDLE TRAY, Material Number 332266 Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous administration of anesthetic fluids.

DateJul 20, 2018
StatusTerminated
ReasonThe catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.
Root causeUnder Investigation by firm
ActionThe firm, B. Braun, sent an "URGENT MEDICAL DEVICE CORRECTION RECALL NOTIFICATION" letters dated 7/20/18 to its customers. The letters described the product, problem and actions to be taken. The letters instruct customers to do the following: 1. Review this Medical Device Correction Notification in its entirety and ensure that all users of the mentioned products in your organization and other concerned persons are informed about this Medical Device Correction Recall Notification. If you are a distributor, please forward this recall notification to your customers. 2. For continued safe use
Quantity2070
DistributionWorldwide Distribution: US (nationwide) and countries of: Australia, Canada, Guatemala and Kuwait.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.