Recall Z-2873-2018 ยท JAK System, X-Ray, Tomography, Computed
GE Healthcare Japan Corporation
GE Healthcare Optima CT660. The systems are intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications. Sold only outside the U.S. under the following product names: CT OPTIMA 660, CT OPTIMA 660 LONG BJG SKD RUS, CT OPTIMA 660 LONG BJG SKD RUSSIA, CT OPTIMA 660 MID BJG SKD RUSSIA, CT OPTIMA 660 SE, CT OPTIMA 660 SE ASIR, CT660 M40 LONG, CT660 M40 MID, CT660 M40 M
| Date | Apr 27, 2018 |
| Status | Terminated |
| Reason | GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry. |
| Root cause | Employee error |
| Action | The firm, GE Healthcare, sent an " URGENT MEDICAL DEVICE CORRECTION" letter dated 4/27/2018 to its Customers on about 04/27/2018. The letter describes the product, problem and actions to be taken. The customers were Instructed to: download and provide the updated Service Manual to the organization that services the equipment on site and informed customers that GE Healthcare would be correcting all affected devices at no cost. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or the local Sales Representative. GE Healthcare expanded the recall to include a |
| Quantity | 92 OUS only |
| Distribution | Worldwide Distribution: US (Nationwide) distribution to states of: AK, AL, CA, CO, CT, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, RI, SC, SD, TN, TX, UT, WI, WV, WY; and countries of:: ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA |