Recall Z-2951-2018 ยท CAC Apparatus, Autotransfusion

Haemonetics Corporation

Haemonetics OrthoPAT System-OPAT QuickConnect Reservoirs. Product Code: OPT-R-1000 The OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs from blood lost intraoperatively and postoperatively during surgical procedures

DateJul 13, 2018
StatusTerminated
ReasonOrthoPAT reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood cannot be processed.
Root causeProcess control
ActionThe firm, Haemonetics, notified customers by an email titled "URGENT: MEDICAL DEVICE RECALL" on July 13, 2018 . The email described the product, problem, and actions to be taken. The customers were instructed to inspect product for defect and return units of the affected lots for replacement with non-affected units; complete the attached customer acknowledgement form via FAX TO +1-781-356-3558 OR SCAN AND E-MAIL TO: CORPORATEREGULATORY@HAEMONETICS.COM. You can contact your local customer service representative at: North America 1.800.537.2802 or email:customerservicena@haemonetics.com.
Quantity348 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122262ORTHOPAT ADVANCE PERIOPERATIVE AUTOTRANSFUSION SYSTEM
Haemonetics Corp., Oct 25, 2012