Recall Z-2891-2018 ยท LLZ System, Image Processing, Radiological

Ortho Kinematics, Inc

Vertebral Motion Analyzer (VMA), 2.3.239 - 2.3.254. Product Usage: VMA software is a quantitative imaging software application intended to be used to process digital image files.

DateMar 2, 2017
StatusOpen, Classified
ReasonOrtho Kinematics notified customers that errors were contained in Radiological Read Report and VMA Report for the VMA test. The error resulted from an administrative error of VMA user account settings and has been fixed.
Root causeSoftware design
ActionOrtho Kinematics notified customers via telephone, email, and letter. Letter dated August 23, 2017 contained the following information: reason for notification, error report information, and requested acknowledgement of receipt.
Quantity1
DistributionU.S. Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133875THE KINEGRAPH VMA (VERTEBRAL MOTION ANALYZER) SOFTWARE VERSION 2.2, THE MOTION NORMALIZER PATIENT HANDLING AND DATA
Ortho Kinematics, Inc., Jan 17, 2014