Recall Z-2837-2018 ยท IYB Tube Mount, X-Ray, Diagnostic

Siemens Medical Solutions USA, Inc

3D TOP Ceiling Stand, model no. 3070039 3074486 3070013 3070021 Product Usage: The ceiling stand is used to hold the X-ray tube and / or the detector and is moveable on a ceiling rail system in three axis. The tube and / or detector can be rotated above a horizontal and a vertical axle. The ceiling stand can be used within a stationary diagnostic X-ray system as well.

DateJun 29, 2018
StatusTerminated
ReasonIn rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break without triggering the safety lock and this could lead to overloading the safety rope which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down during patient positioning. Patients and users may be seriously injured.
Root causeDevice Design
ActionOn 6/29/18, the firm emailed its four affected original equipment manufacturers (OEMs), alerting them to the potential malfunction of the ceiling stand's lifting column that could cause the arm to drop down during patient positioning. Manufacturers were asked to distribute an attached Safety Notice to affected customers. In addition, the firm urged manufacturers to order retrofit kits for all customers and to take any steps needed to ensure that customers install the kits. The kits are free of charge. Manufacturers were also asked to complete a response form and return to the firm. Affected
Quantity699
DistributionUS Nationwide Distribution to accounts in IL, FL, and NY.
FDA recordOpen on fda.gov