ClearanceBrief
Recall Z-2961-2018 ยท JAK System, X-Ray, Tomography, Computed

Bayer Medical Care, Inc.

Overhead Counterpoise Systems (OCS) OCS CONF an accessory device for use with Bayer injector systems as a support for an injector head. , SHORT CEIL MNT, STD HORZ ARM, Catalog # OCS 115AS, Material Number 59880288

DateJun 28, 2018
StatusTerminated
ReasonThe torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.
Root causeEquipment maintenance
ActionOn June 28, 2018, Urgent Medical Device Field Safety Corrective Action letters were distributed to customers. The letters informed the customers of the following: A Bayer Service Representative will be scheduling a service visit to replace the ceiling column assembly of your OCS system(s) at no cost to you.
Quantity1
DistributionThe devices were distributed to the following US states: OH, TX, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K944983MEDRAD COUNTERPOISE SYSTEM
Medrad, Inc., Nov 29, 1994