Recall Z-2878-2018 ยท NGV System, Appendage Closure, Left Atrial

Biosense Webster, Inc.

WaveCrest Left Atrial Appendage (LAA), Occlusion System (32 mm), Catalog Numbers WCR1523 and WCR2523.

DateMay 7, 2018
StatusTerminated
ReasonWhen attempting to recapture the 32-mm device, the tip of the delivery sheath may fold or buckle, resulting in increased retraction force and difficulty or failure to recapture the device.
Root causeComponent design/selection
ActionCoherex Medical notified customers via "URGENT FIELD SAFETY NOTICE Coherex WaveCrest Left Atrial Appendage (LAA) Occlusion System (32 mm) Catalog No: WCR1523, Lot Number: All" letter dated May 7, 2018. The letter provided an overview, and important information about the product description and codes, lots subject to removal, why customers were being contacted, What actions are required/instructions, available assistance/contact information, and Reply Form.
Quantity25 units
DistributionNo U.S. Distribution. International Distribution only.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.