Recall Z-0010-2019 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Radiometer America Inc

ABL90 FLEX analyzers with software versions 3.3MR1 or below. The ABL90 FLEX Analyzer is a portable, automated Analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood. It is intended for use in a laboratory environment, near patient or point-of-care setting.

DateMay 25, 2018
StatusTerminated
ReasonThe ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place: The analyzer is in ready mode Operator lifts the inlet and then closes it again, thereby initiating a rinse Immediately after lifts the inlet and quickly closes it again
Root causeDevice Design
ActionThe firm, Radiometer America, Inc., notified customers via letter dated 05/25/2015. The letter provided the following: Reason for recall, Affect product description and codes, Risk for the patient, What customers must do, Countermeasure provided by Radiometer, contact information and Response Form. The customers were instructed to do the following: 1. Inform users of the ABL90 FLEX analyzer that the inlet is not to be repeatedly lifted and closed. If this does happen, the easiest way to reset the analyzer flow selector manually is to re-install the solution pack. Follow the procedure described
Quantity12,102
DistributionUS Distribution and Internationally to Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120197ABL90 FLEX
Radiometer Medical Aps, Mar 19, 2012