Recall Z-0058-2019 ยท CGN Acid, Folic, Radioimmunoassay

Ortho Clinical Diagnostics Inc

VITROS Immunodiagnostic Products - Folate Reagent Pack(10758750009237) 1513266

DateJul 12, 2018
StatusOpen, Classified
ReasonThe recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.
Root causeDevice Design
ActionOn 12 July 2018, a customer letter was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail to all customers who could process the affected products (i.e., VITROS 3600, 5600 and/or ECi/ECiQ Systems) to inform them of the potential interference of biotin at levels lower than stated in the IFUs and that Ortho is in the process of revising the IFUs for the affected products. Ortho recommended customers follow normal laboratory procedures for troubleshooting of samples containing other assay interferences and it was acceptable to continue u
Quantity7880
DistributionWorldwide Distribution - US Nationwide distribution, and countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, and Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001266VITROS IMMUNODIAGNOSTIC PRODUCTS FOLATE REAGENT PACK 1/2, VITAMIN B12/FOLATE REAGENT PACK 3, PRODUCTS RED CELL FOLATE PA
Ortho-Clinical Diagnostics, Inc., May 2, 2000