Recall Z-3184-2018 ยท CGZ Electrode, Ion-Specific, Chloride
Ortho-Clinical Diagnostics
VITROS Chemistry Products Cl- Slides, Product Code 684 4471 Product Usage: VITROS Cl Slides (Product Code 684 4471) - For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum, plasma and urine using VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Chloride measurements are used in the
| Date | Jul 11, 2018 |
| Status | Terminated |
| Reason | The slides failed to meet current claims for Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantification (LoQ) for urine samples. The measuring range (15-300 mmol/L) for urine samples when tested with the Cl- slide is not affected. The current Cl- LoQ is 5 mmol/L and the LoQ determined by this investigation is 15 mmol/L. |
| Root cause | Under Investigation by firm |
| Action | On 7/11/18, Important Product Correction Notification letters were distributed to customers. The letters instructed customers to perform the following actions: VITROS Cl- Slides and VITROS CREA Slides Instructions for Use will be updated to contain the revised information described in this letter; we will notify you upon availability. In the interim, retain a copy of this customer letter for your records. Calibration Diskette and ADD Data Release Version (DRV) 5992 (and above) will support the revised measuring range for VITROS CREA Slides. We anticipate that DRV 5992 will be available begi |
| Quantity | 168919 |
| Distribution | Worldwide Distribution - US Nationwide- The products were distributed to the following foreign countries: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Nether |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K162020 | VITROS Chemistry Products Cl- Slides Ortho Clinical Diagnostics, Dec 16, 2016 |