Recall Z-2888-2018 ยท ITX Transducer, Ultrasonic, Diagnostic

Civco Medical Instruments Co. Inc.

CIVCO lnfiniti Needle Guide Bracket. used for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers. Product Usage: The bracket and disposable guide provide physicians with a tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers.

DateFeb 23, 2018
StatusTerminated
ReasonA cover failed leak testing during validation testing of new needle guides from a replacement tool.
Root causeDevice Design
ActionTelephone Script to be read: CIVCO is calling to notify our customers regarding a recall action being initiated. CIVCO has recently discovered a dimensional issue in our Infiniti bracket which may allow the needle guide to a puncture a cover. You are being notified, as your facility has been identified as having received product from one of the affected manufacturing lots. Advise on Action to be taken by Customer or Distributor: To return Infiniti needle guide, contact your Account Representative at 800-445-6741 or 319-248-6757. Alternately you may also use email at order@civco.com to sched
Quantity177 boxes
DistributionWorldwide Distribution - US Nationwide in the states of GA, KY, CA, NY , NJ, TX, IL, MA, RI, AZ, and IA and countries of Ukraine, Hong Kong, Italy, Netherlands, Switzerland, United Kingdom, Denmark, Germany, Belgium, Spain, Japan, China, and Canada