Recall Z-2954-2018 ยท BSZ Gas-Machine, Anesthesia

GE Medical Systems China Co., Ltd.

GE Carestation 620/650/650c, A1 and A2 (non-US distribution) anesthesia systems

DateJul 24, 2018
StatusTerminated
ReasonOn systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia.
Root causeNonconforming Material/Component
ActionThe firm issued a medical device correction notice on 07/24/2018. The notice explained the issue and provided safety instructions should the use observe a switch change. The units will be corrected. Questions or concerns may contact 1-800-437-1171 or the local service representative.
Quantity310 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151570Carestation 620/650/650C
Datex-Ohmeda, Inc., Jan 21, 2016