Recall Z-2835-2018 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.

DateMay 17, 2018
StatusOpen, Classified
ReasonAn Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
Root causeUse error
ActionIntuitive Surgical notified customers via a "Field Safety Notice Urgent Medical Device Correction ISIFA2018-06-C da Vinci Xi Miscalibrated Endoscope Controller" letter. The letter provided: Reason for Field Action, Risk to Health, Affected Products (description, part number, serial number), Action taken by Intuitive Surgical, Actions to be taken by the Customer/ User, contact information for Further Information & Support, and an Acknowledgement Form.
Quantity1
DistributionU.S.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131861DA VINCI SURGICAL SYSTEM, ENDOWRIST INSTRUMENTS AND ACCESSORIES
Intuitive Surgical, Inc., Mar 28, 2014