ClearanceBrief
Recall Z-2893-2018 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips Electronics North America Corporation

Philips SureSigns VS3 Vital Signs Monitor, 863069 863070 863071 863072 863073 863074 The SureSigns VS3 monitor is for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility.

DateJul 13, 2018
StatusOpen, Classified
ReasonPhilips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case
Root causeLabeling design
ActionOn July 13, 2018, the firm issued an "URGENT-Medical Device Correction" letter dated July 10, 2018, to affected customers to advise them that that monitors with batteries that have exceeded their specified replacement interval may overheat or ignite. The firm issued a Service Guide addendum included with the letter, which contains information to assist with managing the battery and its replacement. Customers were advised to do the following: Upon receipt of this notification, carefully read the enclosed SureSigns Addendum to the Service Guide. VS Series Customers -Addendum to the Service Gu
QuantityVM 96,324 Units; VS 110,424 Units; VSV 691 Units
DistributionWorldwide distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Canada, Cambodia, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Czech Republic,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163649SureSigns VS3; SureSigns VS4
Philips Medical Systems, Mar 31, 2017