ClearanceBrief
Recall Z-2762-2018 ยท EZL Catheter, Retention Type, Balloon

Degania Silicone, Ltd.

Temperature Sensor Catheter 18FR, Catalogue Number 102201101863BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.

DateJun 21, 2018
StatusTerminated
ReasonThe catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
Root causeEmployee error
ActionUrgent Field Safety Notice letters were distributed to customers on 6/21/18. The letters instructed customers to perform the following: 1. Please forward this information to the immediate attention of the persons responsible for usage of the affected parts. Sub-recall must be initiated by your company to collect any distributed product from the suspected list. 2. Please fill in the attached Recall Response form and indicate the quantity of the affected product in your possession. 3. Promptly return the completed Recall Response form to Degania Silicone by e-mail: Katia@ds-il.com Any affecte
Quantity1025
DistributionWorldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063442ALL SILICONE FOLEY CATHETER, 2-WAY, 3-WAY, WITH TEMPERATURE SENSOR
Degania Silicone , Ltd., Mar 26, 2007