ClearanceBrief
Recall Z-2939-2018 ยท MPB Catheter, Hemodialysis, Non-Implanted

Medical Components, Inc dba MedComp

SOFT-LINE DUO-FLOW DOUBLE LUMEN CATHETER TRAY; T3500 T4000 T4400 T4424

DateJul 12, 2018
StatusTerminated
ReasonThe suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Root causeProcess control
ActionOn July 12, 2018, the firm sent a Product Recall letter to its customers. Customers were provided with a full list of product codes and lot numbers and informed that those items had been recalled due to a potential for the suture wings to crack or break. Customers were informed that implanted catheters with intact suture wings do not need to be removed. However, an alternative method of catheter securement should be initiated for all implanted catheters, due to the potential for the suture wings to crack or break. Customers were instructed to return all unused recalled product to Medcomp. If
DistributionNationwide distribution; worldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K842416DUO FLOW CATHETER-MC 54
Medical Components, Inc., Jul 26, 1984
K983511SOFT-LINE IJ DOUBLE LUMEN CATHETER 11.5FR X 12CM, SOFT-LINE IJ DOUBLE LUMEN CATHETER 11.5FR X 15CM, SOFT-LINE LUMEN CATH
Medical Components, Inc., Jun 8, 1999