ClearanceBrief
Recall Z-2970-2018 ยท HBL Holder, Head, Neurosurgical (Skull Clamp)

Pro-Med Instruments Gmbh

DORO LUCENT Base Unit, REF 1101.021 Headrest System used in open and percutaneous craniotomies and spinal surgeries

DateSep 17, 2018 (initiated Jul 5, 2018)
StatusTerminated
ReasonThe combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the headrest system.
Root causeComponent change control
ActionThe firm initiated the recall by email on 07/05/2018. This stage is only relevant for hospitals / users who are in possession with concerned products of two different revisions (mix-up): a) Discontinue the use of the combination and quarantine the concerned products b) Contact pro med instruments for further instructions relating to temporary replacement possibilities
Quantity30 units
DistributionWorldwide distribution to US states of CA, TX, TN, MS, GA, WV, and OH, Qatar, India, UK, Singapore, Australia, Germany, South Africa, Italy, and Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032331DORO RADIOLUCENT HEADREST SYSTEM AND HORSESHOE HEADRESTS,AND NON-RADIOLUCENT J-ARM RETRACTOR SYSTEM (ALUMINUM ALLOY)
Pro-Med Instruments GmbH, Nov 20, 2003