Recall Z-2925-2018 ยท OLO Orthopedic Stereotaxic Instrument
Mako Surgical Corporation
Stryker Mako Vizadisc Hip Procedure Tracking Kit
| Date | Aug 24, 2018 (initiated Jul 24, 2018) |
| Status | Terminated |
| Reason | Specific lots of the Vizadisc Knee Procedure Tracking Kit and Vizadisc Hip Procedure Tracking Kit have the potential to be damaged on the Vizadisc reflective material causing an inability to be detected by the camera. |
| Root cause | Under Investigation by firm |
| Action | On about 07/24/2018, Stryker sent letters to all their consignees via URGENT MEDICAL DEVICE REMOVAL letter. Instructions for risk mitigation included to perform a visual inspection the discs during pre-surgery setup as the defect can be recognized due to apparent discoloration and in the event that a Vizadisc is damaged, risk may be mitigated by testing the Vizadiscs by attaching them to the probes and arrays during pre-surgery setup, as per Stryker's recommendations in "Surgical Technique Guide." This provides an opportunity for the user to show each instrument to the camera after assembly |
| Quantity | 19,906 total devices |
| Distribution | All devices were distributed in the US in the following states: AK, AL, AR, AZ, CA, CO, CT, FL, IA, ID, KS, KY, LA, MA, MD, MI, MN, MO, NE, NC, NH, NJ, NY, NV, OH, PA, RI, SC, SD, TN. TX, UT, VA, WA, WI. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K121064 | MAKOPLASTY TOTAL HIP APPLICATION, RIO STANDARD SYSTEM Mako Surgical Corporation, Jun 6, 2012 |