Recall Z-2925-2018 ยท OLO Orthopedic Stereotaxic Instrument

Mako Surgical Corporation

Stryker Mako Vizadisc Hip Procedure Tracking Kit

DateAug 24, 2018 (initiated Jul 24, 2018)
StatusTerminated
ReasonSpecific lots of the Vizadisc Knee Procedure Tracking Kit and Vizadisc Hip Procedure Tracking Kit have the potential to be damaged on the Vizadisc reflective material causing an inability to be detected by the camera.
Root causeUnder Investigation by firm
ActionOn about 07/24/2018, Stryker sent letters to all their consignees via URGENT MEDICAL DEVICE REMOVAL letter. Instructions for risk mitigation included to perform a visual inspection the discs during pre-surgery setup as the defect can be recognized due to apparent discoloration and in the event that a Vizadisc is damaged, risk may be mitigated by testing the Vizadiscs by attaching them to the probes and arrays during pre-surgery setup, as per Stryker's recommendations in "Surgical Technique Guide." This provides an opportunity for the user to show each instrument to the camera after assembly
Quantity19,906 total devices
DistributionAll devices were distributed in the US in the following states: AK, AL, AR, AZ, CA, CO, CT, FL, IA, ID, KS, KY, LA, MA, MD, MI, MN, MO, NE, NC, NH, NJ, NY, NV, OH, PA, RI, SC, SD, TN. TX, UT, VA, WA, WI.