Recall Z-2944-2018 ยท JQP Calculator/Data Processing Module, For Clinical Use

Roche Diagnostics Corporation

Roche cobas p 612 pre-analytical system (63x) Model: 07563116001 The cobas p 612 pre-analytical system is a computer controlled fully automatic system for sorting of open and closed barcoded and centrifuged and non-centrifuged sample tubes. The cobas p 612 is intended for use with analyzers that perform tests in the area of clinical chemistry, immuno chemistry, coagulation, hematology, urinalys

DateJul 16, 2018
StatusTerminated
ReasonSample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre-analytical system (63x) (LCP1) with non-filter tips. The possible presence of biological material on the nozzle may cause contamination resulting in potential false positive results, depending on the sensitivity of the analytical technology.
Root causeUnder Investigation by firm
ActionThe firm, Roche, issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated 7/6/18 to its customers on 7/6/18. The letter described the product, problem and actions to be taken. The Customers were instructed to do the following: Complete all sections of the enclosed faxback form and fax it to number 1-866-469-2231 or email it to Roche6840@stericycle.com. " Ensure you have at least a two-week supply of disposable filter tips before the Roche FES arrives to modify your system. " Provide copies of this Urgent Medical Device Correction (UMDC) to other clinicians who may need to be aware of this is
Quantity12 units
DistributionUS Distribution to states of: CA, NY,TX, WA, and WI.
FDA recordOpen on fda.gov