Recall Z-2831-2018 ยท JAC System, X-Ray, Film Marking, Radiographic
Beekley Corporation
S-SPOT REF 777, 777S Adhere to skin for marking and identifying moles in mammography.
| Date | Jun 13, 2018 |
| Status | Terminated |
| Reason | There is a potential for imaging artifact to mimic calcifications on patient's mammograms |
| Root cause | Nonconforming Material/Component |
| Action | On June 14, 2018, Beekley Medical issued an "URGENT MEDICAL DEVICE RECALL" notice dated June 13, 2018 to customers. The notice described the products, problem and actions to be taken. Customers are advised to do the following: - Inspect stock and quarantine affected product - complete and return the acknowledgement response form via fax to: 1-800-735-1234 or mail. - contact firm if inventory of affected product remains with customers at 1-800-233-5539 between 8:00am - 5:30pm EDT. If you have any issues occurred or for further question, contact Quality Assurance & Regulatory Manager at 1-800 |
| Quantity | 3,384,024 pieces in total. |
| Distribution | Worldwide Distribution: US (Nationwide) and countries of; AUSTRALIA, AUSTRIA, BAHRAIN, CANADA, GREAT BRITAIN, ISRAEL, SAUDI ARABIA, SOUTH AFRICA & UNITED ARAB EMIRATES. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.