Recall Z-2830-2018 ยท JAC System, X-Ray, Film Marking, Radiographic

Beekley Corporation

O-SPOT REF 791, 791S Adhere to skin for marking and identifying moles in mammography.

DateJun 13, 2018
StatusTerminated
ReasonThere is a potential for imaging artifact to mimic calcifications on patient's mammograms
Root causeNonconforming Material/Component
ActionOn June 14, 2018, Beekley Medical issued an "URGENT MEDICAL DEVICE RECALL" notice dated June 13, 2018 to customers. The notice described the products, problem and actions to be taken. Customers are advised to do the following: - Inspect stock and quarantine affected product - complete and return the acknowledgement response form via fax to: 1-800-735-1234 or mail. - contact firm if inventory of affected product remains with customers at 1-800-233-5539 between 8:00am - 5:30pm EDT. If you have any issues occurred or for further question, contact Quality Assurance & Regulatory Manager at 1-800
Quantity3,384,024 pieces in total.
DistributionWorldwide Distribution: US (Nationwide) and countries of; AUSTRALIA, AUSTRIA, BAHRAIN, CANADA, GREAT BRITAIN, ISRAEL, SAUDI ARABIA, SOUTH AFRICA & UNITED ARAB EMIRATES.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.