ClearanceBrief
Recall Z-0110-2019 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Medical Systems, LLC

GE Healthcare MRI Systems, Software Version 11.x, sold under the following product names: 1.5T CXK4 LCC MAGNET,1.5T EXCITE, 1.5T EXCITE E/S, 1.5T EXCITE GS, 1.5T EXCITE HYBRID, 1.5T EXCITE TWIN, 1.5T EXCITE UPGRAD, 1.5T HD ES/HS TR W/EXCITE, 1.5T MR MOBILE (SERIAL # R350), 1.5T SIGNA ECHOSPEED, 1.5T SIGNA EXCITE, 1.5T SIGNA EXCITE MRI, 1.5T SIGNA VIBRANT 16CH FIXED, 3.0T MRI TWINSPEED 8CH, 3.0

DateJun 7, 2018
StatusTerminated
ReasonThe latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
Root causeSoftware change control
ActionThe firm, GE Healthcare, notified its customers on or before 06/07/2018 via a letter entitled "URGENT MEDICAL DEVICE CORRECTION". The letter described the product, problem and actions to be taken. The letter informed customers that GE Healthcare will inspect all affected systems and ensure the proper software version is installed. Instructions for customers to confirm the current software included the following steps: 1. Click on tools icon located on the top left of the screen; 2. Click on Service Browser Button; 3. The MR Service Desktop browser will launch (may take 1 minute to launch);
Quantity5,200 total units
DistributionWorldwide Distribution: US (nationwide) including states of:: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV; and countries of: Algeria, Argentina,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0122001.5T INFINITY TWINSPEED MR SYSTEM
GE Medical Systems, Aug 3, 2001
K013636SIGNA INFINITY MR SYSTEM WITH EXCITE TECHNOLOGY
GE Medical Systems, Jan 24, 2002
K030874GE 3.0T SIGNA INFINITY TWINSPEED WITH EXCITE MR SYSTEM
General Electric Co., Mar 28, 2003
K040444GE SIGNA 3.0T WITH EXCITE MR SYSTEM
General Electric Co., Apr 14, 2004
K041476GE SIGNA EXCITE 1.5T MR SYSTEM, GE SIGNA EXCITE 3.0T MR SYSTEM
Ge Medical Systems, LLC, Jun 17, 2004
K052293GE 1.5T AND 3.0T SIGNA HDX MR SYSTEM
Ge Medical Systems, LLC, Sep 21, 2005