Recall Z-0198-2019 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Bovie Medical Corporation

J-Plasma Precise 360, single use, disposable. The Bovie J-Plasma Precise 360 Handpiece is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparascopic surgical procedures.

DateJun 18, 2018
StatusTerminated
ReasonThere is a potential risk associated with the PEEK tip of the J-Plasma Precise 360 Handpiece separating from the handpiece and retained in the tissue.
Root causeNonconforming Material/Component
ActionThe firm, Bovie Medical, notified customers regarding the recall on or about 06/18/2018 via a letter entitled "URGENT: MEDICAL DEVICE RECALL". The letter alerted customers to discontinue to use the device with the affected catalog and lot numbers, to return the impacted handpieces to Bovie Medical, to consider using a device which did not contain the defective component, and how to receive a replacement device. Bovie Medical indicated the resolution to this issue will be a manufacturing and/or design change to the affected handpieces. The letter also stated to complete and return the res
Quantity1,176 total units
DistributionWorldwide Distribution: US (nationwide) to states of:: CA, CO, FL, ID, IL, MD, MO, NJ, NY, OH, OR, PA,, TX.; and countries of: Austria, Bahrain, Belgium, Bulgaria, Cyprus, Finland, Great Britain, Hungary, Ireland, Israel, Italy, Qatar, Scandinavia, Slovakia, Switzerland, Turkey, and United Arab Emi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152570Bovie J-Plasma Precise 360 Handpiece
Bovie Medical Corporation, Nov 2, 2015