Recall Z-2597-2018 ยท FKX System, Peritoneal, Automatic Delivery

Baxter Healthcare Corporation

power cord for: (a) AMIA Automated PD Cycler, Product Code: 5C9320; (b) Kaguya PD System, Product Code: T5C8500 (Japanese distribution only)

DateJul 12, 2018
StatusTerminated
ReasonThere were instances where the power cord socket detached from the back of the AMIA cycler when the power cord was unplugged from the socket.
Root causeUnder Investigation by firm
ActionMethod of Notification: A Safety Alert communication will be sent to affected customers via U.S.P.S., first class mail. Mitigating Instructions: Baxter is asking customers to: 1. Unplug the devices from the wall before unplugging the power cord from the back of the cycler. 2. Confirm the receipt of the letter by completing the enclosed customer reply form and returning it to Baxter by either mail, fax, or scanned email.
Quantity5445 units
DistributionUS and Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151525Amia Automated PD System with Sharesource
Deka Research and Development, Oct 9, 2015