Recall Z-2537-2018 ยท JAI Couch, Radiation Therapy, Powered

Forte Automation Systems Inc

Patient Positioning System with KRC2 controller using software versions 2.3.1 2.3.10 Product Usage: The patient positioning system is a SCARA designed robotic arm designed to position a patient for medical procedures prescribed by oncologists and others that require a high degree of accuracy and repeatability.

DateSep 5, 2017
StatusTerminated
ReasonCommunications between the Patient Positioning System and the accuracy filter can periodically fail with no clear indication to the operator.
Root causeSoftware design
ActionOn September 5, 2017 Forte has distributed an Urgent Medical Device Correction letter with the following information, Subject: Robotic Treatment Table Positioning Error Commercial Name of Affected Product: Patient Positioning System Affected Version(s) / Lot(s): KRC2 Controllers with software version 2.3.1 2.3.10 Reference / FSCA Identifier: 3003900998-2017-00001 Type of Action: Notification and Correction This letter provides users with a Description of the Problem, Details, Recommended User Action and Forte Automation Systems, Inc. Actions. This document also Pending resolution with a perm
Quantity13
DistributionUS in the state of California
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122413PATIENT POSITIONING SYSTEM
Forte Automation Systems, Inc., Sep 13, 2012