Recall Z-2604-2018 ยท MNI Orthosis, Spinal Pedicle Fixation

New Era Orthopaedics, LLc

Modera Modular Pedicle Screw System Shank Insertion Instrument, 3-107-07

DateFeb 13, 2017
StatusTerminated
ReasonThis is a retrospective recall of an instrument used in spine fusion procedures with Modera Modular Pedicle Screw system implants. The firm received a customer complaint in which excessive forces were applied to the shank insertion instrument and the laser weld failed.
Root causeDevice Design
ActionOn February 13, 2017, the firm emailed its distributors to notify them that their Rev. A devices would be replaced with Rev. B devices, and asked the distributors to mail back the Rev. A devices.
Quantity6 pieces