Recall Z-2525-2018 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Siemens Medical Solutions USA, Inc

AXIOM Artis with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic imaging and interventional procedures.

DateMay 25, 2018
StatusTerminated
ReasonDue to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure
Root causeProcess control
ActionSiemens issued on May 25, 2018 an URGENT: MEDICAL DEVICE CORRECTION customer notification to affected customers and advised a hardware update will be initiated . Users advised of the problem, potential risks and action to be taken. Siemens will correct the error by replacing all affected cooling units with Update Instruction AX076/17/S. Users may continue to use the Artis system while awaiting implementation of the corrective action.
Quantity6 units US
DistributionHI , IL, MI, NC, NY, VA, WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021021MODULAR ANGIOGRAPHY SYSTEM AXIOM ARTIS
Siemens Medical Solutions USA, Inc., Jun 6, 2002