Recall Z-2574-2018 ยท GNP Antiserum, Cf, Epstein-Barr Virus

Diamedix Corporation

Diamedix Is-EBV-EA-D IgG Enzyme Immunoassay Test Kit for In Vitro Diagnostic Use Catalog No. 720-640

DateJan 5, 2018
StatusTerminated
ReasonContains a gel-like contaminant that may affect product performance.
Root causeUnder Investigation by firm
ActionDiamedix sent a letter dated January 5th, 2018 was sent to all their customer with the following: Products contains a gel-like contaminant that may affect product performance.An inspection of the product in the Diamedix inventory confirmed this observation. Although, no failed runs have been directly attributed this issue, we feel that the best course of action is to replace the affected diluent. If you are using the reagent lot specified above, please discontinue use immediately. In order to process your replacement, please complete the attached Certificate of Destruction and return this do
Quantity147
DistributionAZ, CA, FL, IL, IN, KY, MA, MI, MN, MO, NH, NJ, NY, OH, OK, PA, RI, TN, TX, and WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K981831IS EBV-EA-D IGG ELISA TEST SYSTEM
Diamedix Corp., Feb 16, 1999