ClearanceBrief
Recall Z-2511-2018 ยท GDW Staple, Implantable

COVIDIEN MEDTRONIC

Endo GIA 45 mm Articulating Vascular/Medium Loading Unit with Tri-Staple Technology, Item Code EGIA454AVM

DateJul 19, 2018 (initiated May 22, 2018)
StatusOpen, Classified
ReasonDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Root causeProcess control
ActionOn May 22, 2018, all U.S. consignees received an UPDATED URGENT MEDICAL DEVICE RECALL and RETURN FORM notice via Federal Express or certified. The initial notice was sent to impacted customers on April 26, 2018. Customers are advised to take the following required actions: 1. Immediately discontinue use, and quarantine affected products listed 2. Return affected product as indicated. 3. Complete the Recalled Product Return Form even if you do not have inventory and submit by either method below: * Product purchased directly from Medtronic: feedback.customerservice@Covidien.com or fax to (80
DistributionNationwide U.S.A. Internationally to: Albania, Algeria, Austria, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Iceland, Ireland, Israel, Italy, Ivory Coast, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Mal
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.