ClearanceBrief
Recall Z-2593-2018 ยท GDJ Clamp, Surgical, General & Plastic Surgery

Integra LifeSciences Corp.

MAYFIELD Infinity XR2 Base Unit-Extended, Part Number A2079E The MAYFIELD Infinity XR2 Base Unit is intended to be used to support a patient during diagnostic examination and/or surgical procedures where a rigid support between surgical table and headrest, or skull clamp is necessary, positional freedom is required and where X-ray imaging modalities will be used.

DateJun 29, 2018
StatusTerminated
ReasonThere is a potential for the knob assembly in the swivel adapter to fracture/break during use.
Root causeDevice Design
ActionThe firm, Integra, sent an "Urgent Voluntary Medical Device Correction Field Safety Notice" dated 6/29/18 to customers. The notification letter instructed customers to perform the following: 1) Review your MAYFIELD XR2 instructions for use. Please pay particular attention to the CAUTIONS listed on page 5 of the accompanying instructions for use section as noted below. " Always make sure the base unit is properly secured to the operating table " The base unit must not be used if the device appears to be damaged or functioning incorrectly " Over-tightening the mechanisms adjustment screws ma
Quantity17 units
DistributionWorldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV; and to countries of:: Australia, Canada, China, Dominican Re
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.