Recall Z-2441-2018 ยท RGW Uv/Ir Research

Zeiss, Carl Inc

Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope

DateMay 29, 2018
StatusTerminated
ReasonUnder certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.
Root causeRadiation Control for Health and Safety Act
ActionThe Company will immediately notify by direct contact (when available) and/or express courier each first line purchaser and/or known user site with informing them of the risk, how to prevent it and the notice of action. The Company will dispatch factory trained service representatives to perform following actions: o Installation of Hot Fix # 14 (or newer) for ZEN 2.3 o Installation of Firmware 1 .209 (or newer) on nodes x l 5 and x l 9 (Main Controller 2)
Quantity8
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.