ClearanceBrief
Recall Z-2544-2018 ยท NDS Media, Culture, Ex Vivo, Tissue And Cell

Life Technologies, Corp.

Gibco Dulbecco's Modified Eagle Medium (DMEM), high glucose, pyruvate, Model Number 11995115. Tissue and cell culture medium.

DateJul 25, 2018 (initiated May 10, 2018)
StatusTerminated
ReasonIt has been determined that the fill port tube seal integrity was compromised during the manufacture of the product lot. This may cause media to leak out of the primary sterile packaging into outer non-sterile dust cover packaging, and subsequently mold may grow in the pooled, leaked media.
Root causeUnknown/Undetermined by firm
ActionUrgent Medical Device Recall Notification letters dated 6/28/18 were distributed to customers to notify them of the recall of GIBCO CTS AIM V Serum-Free Media (SFM); Gibco Dulbecco's Modified Eagle Medium (DMEM), high glucose, pyruvate; Gibco Dulbecco's Modified Eagle Medium (DMEM), high glucose; and Gibco RPMI 1640 Medium. The letters instructed customers to do the following: For the above mentioned lots of product, please complete an inspection of the seal integrity of your inventory. There should not be visual evidence of media within the space between the media bag itself and the dust c
Quantity45
DistributionWorldwide distribution. US states: CA, CO, CT, FL, GA, MA, MD, PA, RI, UT, and WA., Canada, France, Great Britain, Japan, Korea, Netherlands, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.