Recall Z-2599-2018 ยท LTT Panels, Test, Susceptibility, Antimicrobial

Beckman Coulter Inc.

MicroScan Pos Combo Panel Type 43, Catalog number B1017-217for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram-positive cocci and Listeria monocytogenes.

DateMay 4, 2018
StatusTerminated
ReasonThe drug dilution sequence for Moxifloxacin printed on the panel is incorrect. As a result, incorrect MIC results for Mxf could be reported.
Root causeLabeling design
ActionA recall notification letter was sent the week of 05/01/2018 via email, mail, and phone if needed.
Quantity686 boxes
DistributionNJ, TX, IA, PA, FL, NE, OH, NY, CA, MO, OK, LA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020160DRIED GRAM-POSITIVE MIC/COMBO PANELS WITH AMPICILLIN
Dade Behring, Inc., Mar 14, 2002