Recall Z-2792-2018 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Megadyne Medical Products, Inc.

MegaPower Electrosurgical Generator, Product code 1000, GTIN 10614559103715

DateAug 13, 2018 (initiated May 15, 2018)
StatusTerminated
ReasonIf two devices are plugged into the same channel/port in the generator, both devices activate. If one device is left laying on the patient, the result could be burning of the patient.
Root causeUse error
ActionThe firm sent letters in June 2018, informing their customers of the issue. The firm will send labels that can be applied to the product to remind them not to plug 2 active devices into the same channel at a time. Updated 08/13/19 Starting on 06/29/18, updated Medical Device Correction Notifications and labels were mailed via second day service. Customers were asked to complete and return Business Reply Forms.
Quantity8380 units
DistributionWorldwide distribution. US Nationwide, Mexico, Europe, Middle East, Latin America, Africa, and Asia Pacific Markets
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050579MEGA POWER ELECTROSURGICAL GENERATOR
Megadyne Medical Products, Inc., Mar 24, 2005