Recall Z-2410-2018 ยท KDP Regulator, Vacuum

Ohio Medical Corporation

Vacuum Regulator, Push-T-Set Intermittent Suction Unit (PTS-ISU) Product Usage: The vacuum regulator is intended to be used in a medical facility as a means to evacuate media (i.e. fluids) from the body.

DateMar 12, 2018
StatusTerminated
ReasonFinal quality control testing was not completed before the devices were distributed..
Root causeEmployee error
ActionOhio Medical notified customers via phone call on about 03/12/2018. The notification included identification of the devices distributed to that customer, explanation that due to the need to the missed additional testing they were removing the devices, and providing a replacement until the testing can be completed.
Quantity11 units
DistributionUS Nationwide Distribution in the states of NC, IN, TN, LA,, FL MI, IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.